简单介绍:
GFC 17 Mass Flow Controller P/N:GFC37S - BADL2 - C0A。 气体:氢气。 *大流量:80.00升/分钟。 外壳材料:不锈钢。 密封:氟橡胶,阿®(标准). 配件:1 / 4英寸压缩。 显示:液晶读数显示。 电源输入:12伏直流。 输入/输出信号:0-5伏/ 0-5伏
详情介绍:
GFC 17 Mass Flow Controller
P/N:GFC37S - BADL2 - C0A。
气体:氢气。
*大流量:80.00升/分钟。
外壳材料:不锈钢。
密封:氟橡胶,阿®(标准).
配件:1 / 4英寸压缩。
显示:液晶读数显示。
电源输入:12伏直流。
输入/输出信号:0-5伏/ 0-5伏

Model GFC thermal Mass Flow Controllers are designed to indicate and control set flow rates of gases.

The GFC combines the characteristics, and accuracy of conventional mass flow devices into a unique compact design at low costs previously unattainable. Each of these controllers incorporates an advanced straight tube sensor in conjunction with flow passage elements constructed of aluminum and brass for non-corrosive gases or 316 stainless steel for corrosive applications. Zero and span adjustments are accessible from the outside of transmitters.

Principles of Operation

Metered gases are divided into two laminar flow paths, one through the primary flow conduit, and the other through a capillary sensor tube. Both flow conduits are designed to ensure laminar flows and therefore the ratio of their flow rates is constant.

Two precision temperature sensing windings on the sensor tube are heated, and when flow takes place, gas carries heat from the upstream to the downstream windings. The resultant temperature differential is proportional to the change in resistance of the sensor windings.

A Wheatstone bridge design is used to monitor the temperature dependent resistance gradient on the sensor windings which is linearly proportional to the instantaneous rate of flow.

Output signals of 0 to 5Vdc and 4 to 20mA are generated indicating mass molecular based flow rates of the metered gas.

The combined gas streams flow through a proportionating electromagnetic valve with an appropriately selected orifice. The closed loop control circuit continuously monitors the mass flow output and maintains it at the set flow rate.

Flow rates are unaffected by temperature and pressure variations within stated limitations.

General Description

Compact, self contained GFC mass flow controllers are designed to indicate and control flow rates of gases. The rugged design coupled with instrumentation grade accuracy provides versatile and economical means of flow control. Aluminum or stainless steel models with readout options of either engineering units (standard) or 0 to 100 percent displays are available.

The built-in electromagnetic valve allows the flow to be set to any desired flow rate within the range of the particular model. The valve is normally closed as a safety feature to ensure that gas flow is shut off in case of a power outage. Setpoints are controlled either locally or remotely.

The LCD readout built into the top of the transducer is tiltable over 90 degrees to provide optimal reading comfort. It is connected to the transducer by a standard modular plug, and is readily removable for remote reading installations. Transducers without LCD readout are offered for OEM applications.

GFC mass flow controllers are available with flow ranges from 10 sccm to 1000 sL/min N2. Gases are connected by means of 1/4”, 3/8”, or optional 1/8” compression fittings and 3/4” FNPT fittings. Optional fittings are available. These controllers may be used as benchtop units or mounted by means of screws in the base.

Transducer power supply ports are fuse and polarity protected.

Leak Integrity

1 x 10-7 smL/sec of helium maximium to the outside environment.

Specification Information

Gas:                                                          Hydrogen
Maximum Flow Rate:
                               80.00 std. L/min

Housing Material:                                     Stainless Steel
Seals:                                                      
Viton-A® (Standard)
Fittings:                                                    1/4 inch compression
Display:                                                    LCD readout display
Power Input:                                            12 VDC
Input/Output Signal:                                 0-5 VDC / 0-5 VDC


ACCURACY (including linearity):

±1.5% of full scale, including linearity for gas temperatures of 59 FF to 77 FF (15 FC to 25 FC) and pressures of 5 to 60 psia (0.4 to 4.1 bars). Optional ±1% of full scale (certified calibration accuracy) associated with a given set of temperature and pressure values.

REPEATABILITY:
±0.5% of full scale.

RESPONSE TIME:
Generally 2 seconds to within ±2% of actual flow rate over 25 to 100% of full scale.

TEMPERATURE COEFFICIENT:
0.15% of full scale / FC.

PRESSURE COEFFICIENT:
0.01% of full scale / psi (0.07 bar).

PRESSURE DROP:
See table 5.

OPTIMUM GAS PRESSURE:
25 psig (1.73 bars).

MAX GAS PRESSURE:
1000 psig (70 bars) maximum GFC 17, 37, 47. 500 psig (34.5 bars) GFC 57, 67, 77.

MAX DIFF. PRESSURE:
GFC 17, 37, 57, 67, and 77 50 psi (3.4 bars), GFC 47, 40 psi (2.7 bars).

GAS and AMBIENT TEMP:
41 FF to 122 FF (5 FC to 50 FC).

**MATERIALS FLUID CONTACT:
a. Aluminum models GFC Series: anodized aluminum, 316 stainless steel, brass and Viton7 O-rings.
b. Stainless Steel models GFC17S, 37S, 47S, 57S, 67S and 77S: 316 stainless steel and Viton7 O-rings.
Optional O-rings: Buna, EPR and Kalrez7.

ATTITUDE SENSITIVITY:
No greater than +15 degree rotation from horizontal to vertical; standard calibration is in horizontal position.

OUTPUT SIGNALS:
Linear 0-5 Vdc. (1000 ohms min. load impedance); 4-20 mA (0-500 ohms loop resistance) Max noise ±20mV.

COMMAND SIGNALS:
Analog 0-5 Vdc or 4-20 mA for remote set point mode; NPN compatible purge /valve off.

CONNECTIONS:
GFC 17 and 37: 1/4” compression fittings. Optional: 1/4” VCR7, 1/8”or 3/8 compression fittings (GFC17).
GFC 47: 3/8” compression fittings.
GFC 57: 3/8” compression fittings.
GFC 67: 1/2” compression fittings.
GFC 77: 3/4” FNPT fittings. Optional: 3/4” compression fittings (GFC77).

LEAK INTEGRITY:
1 x 10-7 smL/sec of helium maximum to the outside environment.

TRANSDUCER INPUT POWER:
+12 Vdc, 800 mA; +24 Vdc, 650 mA optional.

CIRCUIT PROTECTION:
Circuit boards have built-in polarity reversal protection. Resettable fuses provide power input protection.

DISPLAY:
3-1/2 digit LCD, 0.5” high characters.

CE COMPLIANT:
EN 55011 class 1, class B; EN50082-1.

H-Tech Shipping Information
Estimated Ship Dimensions: 8" x 8" x 6"
• Estimated Ship Weight: 4 lbs
• Lead time: 16 days
• Shipping origin: Orangeburg, United States